Home > Sterilizers > Hospital Autoclave (Steam Sterilizer)
Hospital Autoclave (Steam Sterilizer)
Hospital Autoclave (Steam Sterilizer)

Hospital Autoclave (Steam Sterilizer)

MADDA/MASDA

MADDA and MASDA are hospital steam sterilizers built for central sterile services departments (CSSD). The range covers 15 models from 90 L to 845 L (1–12 STU), in single-door (MASDA) and double-door pass-through (MADDA) versions, with an EN 285 compliant sterilization system and a 316L stainless steel chamber. An integrated steam generator and HMI touch panel deliver reliable saturated-steam sterilization of heat- and moisture-resistant medical instruments.

  • EN 285
  • ISO 13485
  • 316L chamber
  • Single / double door
  • 90–845 L chamber volume, 1–12 STU
  • Single-door or double-door (pass-through) configuration
  • Integrated steam generator and HMI touch control
  • Pre- and post-vacuum, automatic leak test
  • Up to 30% less water with the optional recovery system

Single-Door or Double-Door Steam Sterilizer?

Both series share the same chamber volumes and steam generators; the difference is the door design and the installation. Your CSSD layout decides.

Single-door autoclave layout: loads go in and come out on the same side

8 models · 90–845 L

MASDA single-door autoclave

Loading and unloading happen in the same room. Suited to operating theatre preparation areas and compact sterile processing units.

  • No wall opening required
  • Installed within one room
  • Compact model from 90 L
Double-door autoclave layout: loading on the dirty side, unloading on the clean side

7 models · 165–845 L

MADDA double-door autoclave

Mounted in the wall between the dirty and clean zones. Loads go in on the dirty side and come out on the sterile side, keeping the CSSD workflow one-directional.

  • Dirty/clean zone separation (pass-through)
  • Built for central sterile services departments
  • 7 models from 165 L to 845 L

Models and Capacities

Size your sterilizer by daily tray volume, container format (STU 300×300×600 mm or ISO 600×400×200 mm) and available installation space.

Models by capacity

Models by capacity
Chamber volume (L)Chamber W×H×D (mm)Generator (kW)STU containersISO containersSingle doorDouble door
90 370×370×665 15 1 0 MASDA 0090 —
165 370×670×665 30 2 2 MASDA 0165 MADDA 0165
250 500×500×1005 30 1 1 MASDA 0250 MADDA 0250
305 670×670×665 40 4 2 MASDA 0305 MADDA 0305
435 670×670×955 40 6 4 MASDA 0435 MADDA 0435
565 670×670×1245 50 8 6 MASDA 0565 MADDA 0565
710 670×670×1580 60 10 8 MASDA 0710 MADDA 0710
845 670×670×1880 75 12 8 MASDA 0845 MADDA 0845

Scroll the table sideways →

W: width, H: height, D: depth. STU: 300×300×600 mm, ISO: 600×400×200 mm container.

All models: technical dimensions

All models: technical dimensions
 Chamber (mm)Outer dimensions (mm)Steam generatorContainer quantity
ModelVolume (L)WHDWHDPower (kW)Volume (L)STUISO
MASDA 0090 90 370 370 665 770 1860 1205 15 40 1 0
MASDA 0165 165 370 670 665 770 1860 1400 30 40 2 2
MASDA 0250 250 500 500 1005 890 1860 1590 30 40 1 1
MASDA 0305 305 670 670 665 1050 1860 1300 40 60 4 2
MASDA 0435 435 670 670 955 1050 1860 1600 40 60 6 4
MASDA 0565 565 670 670 1245 1050 1860 1700 50 60 8 6
MASDA 0710 710 670 670 1580 1050 1860 2250 60 80 10 8
MASDA 0845 845 670 670 1880 1050 1860 2400 75 80 12 8
MADDA 0165 165 370 670 665 1025 1860 1050 30 40 2 2
MADDA 0250 250 500 500 1005 900 1860 1600 30 40 1 1
MADDA 0305 305 670 670 665 1350 1860 1250 40 60 4 2
MADDA 0435 435 670 670 955 1350 1860 1440 40 60 6 4
MADDA 0565 565 670 670 1245 1350 1860 1585 50 60 8 6
MADDA 0710 710 670 670 1580 1350 1860 2160 60 80 10 8
MADDA 0845 845 670 670 1880 1350 1860 2320 75 80 12 8

Scroll the table sideways →

MASDA: single door, MADDA: double door (pass-through). All dimensions in millimetres.

Technical Features

Construction and vacuum system

  • 316L stainless steel chamber, steam jacket and generator
  • 304 stainless steel body
  • Integrated steam generator
  • Pre-vacuum and post-vacuum technology
  • High-capacity, low-noise liquid ring vacuum pump

Sterilization programs

  • Wrapped loads 134 °C
  • Wrapped loads 121 °C
  • Fast cycle 134 °C
  • Heavy surgical instrument cycle
  • Bowie-Dick test and automatic leak test
  • Custom material programs

Safety and alarms

Pressure, temperature, vacuum and door position are monitored throughout the cycle.

  • Automatic door locking and pressure-controlled door safety
  • Emergency stop button
  • Password-protected user access
  • Audible and visual alarms
  • HEPA-filtered air inlet

HMI control panel

  • Touch-screen interface
  • Real-time cycle monitoring
  • Programmable sterilization cycles
  • User-based access management
  • Alarm and warning notifications

Energy and water savings

The optional water recovery system recycles vacuum pump process water through a closed-loop cooling circuit.

  • Up to 30% lower water consumption
  • Eco mode water recovery
  • Energy-saving standby mode
  • Automatic early start

Standards and Certificates

The sterilization system complies with EN 285; the unit is manufactured in an ISO 13485 certified facility.

  • EN 285 Large steam sterilizers
  • ISO 13485 Medical device quality management
  • ISO 9001:2015 Quality management system
  • ISO 14001:2015 Environmental management system
  • ISO 45001:2018 Occupational health and safety
  • ISO 10002:2018 Customer satisfaction management
  • GMP Good manufacturing practice

Applications

From a compact 90 L unit to an 845 L system, the range fits healthcare facilities of every size.

  • Hospitals and CSSDs
  • Operating theatres
  • Intensive care units
  • Endoscopy centers
  • Clinics and health centers
  • University and research laboratories
  • Microbiology laboratories
  • Veterinary facilities
  • Industrial sterilization

CSSD Planning from One Manufacturer

Mixta plans the hospital steam sterilizer together with the whole CSSD. Washer-disinfectors, packing tables, sterilization baskets and sterile transfer units come from a single manufacturer in Ankara, Türkiye.

Our project planning and design team works with you on the CSSD layout across dirty, clean and sterile zones.

  • 2007Founded
  • 110+Countries
  • 6Continents

Frequently Asked Questions

What is the difference between MADDA and MASDA?

MASDA is single-door; MADDA is double-door (pass-through). Chamber volumes and steam generators are the same; the difference is the door design and the installation.

Which capacities are available?

Single-door MASDA starts at 90 L and double-door MADDA at 165 L; both go up to 845 L, a capacity of 1–12 STU (300×300×600 mm).

How is a hospital steam sterilizer different from a tabletop autoclave?

Tabletop autoclaves are small clinical units (EN 13060). A hospital steam sterilizer is built for high-volume, continuous CSSD use; large steam sterilizers of 60 L and above fall under EN 285. For small volumes, see our tabletop autoclave.

When do I need a double-door autoclave?

When the CSSD separates dirty and clean zones with a wall. Loads enter on the dirty side and leave on the sterile side, so the workflow stays one-directional.

Which sterilization programs are included?

134 °C and 121 °C wrapped cycles, a 134 °C fast cycle, a heavy surgical instrument cycle, Bowie-Dick test, automatic leak test and custom material programs.

How much water does it use?

Optimized vacuum and steam algorithms plus the optional water recovery system cut water consumption by up to 30%.

What determines the price?

Chamber volume, single or double door, optional equipment (for example water recovery) and project scope. Submit your project with capacity and installation details for a quote.

Related Products

Plasma Sterilizers

Plasma Sterilizers MPS Series

Steam Sterilizers

Steam Sterilizers MBS Series

Autoclave (23 L)

Autoclave (23 L) MFO 8003

Customer ServiceMixta Medical

Hello!

Can I help you?

04:27

Cookie Preferences & Privacy Management